MEDUVENT Standard
More flexibility from turbine technology

Wide variety of potential applications
Long distances or transport times? No power or oxygen supply? With a ventilator that maintains ventilation for an average of 8 hours*, even without an external gas supply, you don’t need to worry about the technology. MEDUVENT Standard is one of the smallest turbine-driven ventilators in the world and makes you independent of oxygen cylinders. If required, however, you can still administer additional oxygen to the person undergoing treatment at any time via the universally compatible oxygen inlet tube. This allows inspiratory oxygen concentrations of 21 % to 100 % to be achieved.
In emergency response, the speed at which you initiate therapy on that person is often of key importance for the success of their therapy. MEDUVENT Standard enables you to ventilate ergonomically in a controlled and efficient manner - without having to set ventilation parameters by a complicated process. As a result, the patient is taken care of as quickly as possible, safely, and in accordance with guidelines.
* assuming typical ventilation settings for an adult patient
Five benefits of MEDUVENT Standard
- Protected by an alarm system, automatic function check, and hygiene filter
- Loss-free oxygen supply from 21 % O₂ to 100 % O₂
- With a weight of just 2.1 kg and a volume of 3.5 l, one of the smallest, lightest emergency medicine and transport ventilators in the world
- Innovative turbine technology: Works with and without an oxygen supply
- Long operating time: Average of 8 hours for one battery charge (assuming typical ventilation settings for an adult)
For example, we had a patient whose state of health improved significantly after she had been ventilated with MEDUVENT Standard.
Dr. Annet Alenyo Emergency Physician
More product information
Options
- Integrated ventilation modes
- IPPV (Intermittent Positive Pressure Ventilation): Volume-controlled mandatory ventilation
- CPAP (Continuous Positive Airway Pressure)
- manual
- Other available modes and functions
- S-IPPV (Synchronized Intermittent Positive Pressure Ventilation): Assisted volume-controlled ventilation with a trigger time slot of 100 % of expiratory time. For volume-controlled ventilation with variable mandatory minute volume. Provides a mandatory mechanical breath for every attempt at spontaneous breathing.
- SIMV (Synchronized Intermittent Mandatory Ventilation): Assisted volume-controlled ventilation with a trigger time slot of 20 % of expiratory time
- CPAP + PS (Continuous Positive Airway Pressure + Pressure Support) For pressure support if spontaneous breathing is insufficient or exhausted.
- PRVC + PS (Pressure-Regulated Volume-Controlled ventilation + Pressure Support) Combines the benefits of pressure-controlled and volume-controlled ventilation. Pressure support PS can be used for patients who breathe spontaneously from time to time.
Accessories & portable units
If you are still looking for the right accessories, then take a look at the catalog for our scope of supply! It contains the matching WEINMANN Emergency accessories and consumables for your MEDUVENT Standard.
Matching portable units
- LIFE-BASE light XS
- LIFE-BASE 1 NG XS
- LIFE-BASE 1 NG XL
- LIFE-BASE 3 NG
Download area
If you are looking for brochures, images, videos, PC simulations, instructions for use or certificates, you can find all product information in our Download area.
LEARN MORE
Technical details
- Dimensions (W x H x D):
206 mm x 137 mm x 130 mm
- Weight, incl. battery:
2.1 kg
- Product class according to Directive 93/42/EEC:
IIb
- Temperature (operating conditions):
-20 °C to +50 °C
- Humidity (operating conditions):
5 % rh to 95 % rh, no condensation
- Air pressure (operating conditions):
540 hPa to 1,100 hPa
- Altitude above mean sea level (operating conditions):
-500 m to 5,000 m
- Operating time (lithium ion battery):
average 8 hours
- Charging time (0 % – 95 %, lithium ion battery):
2.5 h (assuming settings for an adult)
- Display:
TFT color display 4.3"
- Data storage:
internal device memory and SD card
- Ventilation modes (mandatory/triggered) volume-controlled:
- IPPV
- S-IPPV
- SIMV
- manual
- PRVC + PS
- Ventilation modes (spontaneous):
CPAP
CPAP + PS
- Operating gas:
Ambient air; medical oxygen (100 % O₂) or concentrator oxygen (93 % O₂) can also be supplied
- Operating pressure range:
0.3 bar to 6 bar
- Monitoring:
- measured values displayed: MVi, O₂i
- Bar graph: pressure with pAw
- Maximum outlet flow:
150 l/min
- Tidal volume:
50 ml to 2,000 ml
- Ventilation rate:
5/min to 40/min
- PEEP:
0 mbar to 20 mbar
- Pressure limit (pMax):
10 mbar to 65 mbar
- I:E:
1:2 fixed / 1:1 in manual mode
- Standards used:
- Consistent serviceability confirmed on the ground and in the air to EN 13718-1, RTCA/DO 160G, EN 60601-1-12
- Ruggedness confirmed by successful shock and vibration testing to MIL-STD 810 G, RTCS/DO 160G and EN 1789
- Emergency and transport ventilator confirmed to EN 794-3 and ISO 10651-3

Ideal monitor layout
Displays the most important measurements and settings
User-friendly operation
Navigation knob is quick and easy to operate
Li-ion rechargeable battery
Average battery operating time of 8 hours (with typical ventilation settings for an adult patient)
Data storage and updates
SD memory card makes it easy to transfer device configurations and to perform software updates yourself
Connection for O₂ supply
Via O₂ cylinders, O₂ concentrators and gas supply systems
Connection for measuring circuit
Enables pressure to be measured and PEEP to be controlled, as well as oxygen to be routed to the patient
Accessories connection accessible from the front
For MEDUtrigger
Connection for ventilation hose
Connects the device to the breathing circuit
Hygiene filter
Protects the device from viral and bacterial contamination
Adapt oxygen supply flexibly

If you want to supply your patients with 100 % oxygen and then adapt the dose, simply connect the inhalation outlet of the pressure reducer to the oxygen inlet of MEDUVENT Standard using the globally-compatible oxygen inlet tube. You can then set the desired flow.
As MEDUVENT Standard runs purely on electricity, and not on compressed gas, the device does not use any oxygen to run. If you do not have any medical oxygen available but still want to ventilate with more than 21 % oxygen, you can use concentrator oxygen. To do this, just change the operating gas and connect the concentrator or the oxygen cylinder to MEDUVENT Standard.
It goes without saying that you can also operate MEDUVENT Standard with ambient air, without any oxygen source at all. The hygiene filter cleans the air used, removing viruses, bacteria, and particles, thus protecting both people and the device from contamination.
Training courses and interactive training
Our training video gives you an initial impression of MEDUVENT Standard and explains how to operate it.
If you want to find out what it is like to operate one of the smallest turbine-driven ventilators in the world, you are welcome to use our interactive simulation software. You can use it to try out all the menus and functions of MEDUVENT Standard.