MEDUVENT Standard
Ventilate Wherever You Go

Wide variety of potential applications
Long distances or transport times? No power or oxygen supply? With a ventilator that maintains ventilation for 7.5 hours*, even without an external supply of compressed gas, you don’t need to worry about the technology. MEDUVENT Standard is one of the smallest turbine-driven ventilators in the world and makes you independent of oxygen cylinders. If required, however, you can still administer additional oxygen to the person undergoing treatment at any time via the universally compatible oxygen inlet tube. This allows inspiratory oxygen concentrations of 21 % to 100 % to be achieved.
In emergency response, the speed at which you initiate therapy on that person is often of key importance for the success of their therapy. MEDUVENT Standard enables you to ventilate ergonomically in a controlled and efficient manner - without having to set ventilation parameters by a complicated process. As a result, the patient is taken care of as quickly as possible, safely, and in accordance with guidelines.
* assuming typical ventilation settings for an adult patient
Six benefits of MEDUVENT Standard
- Loss-free oxygen supply from 21 % O₂ to 100 % O₂
- Pressure and flow curves displayed for clear monitoring
- Treatment by means of volume-controlled and pressure-controlled ventilation
- With a weight of just 2.1 kg and a volume of 3.5 l, one of the smallest, lightest emergency medicine and transport ventilators in the world
- Innovative turbine technology: Works with and without an oxygen supply
- Protected by an alarm system, automatic function check, and hygiene filter
More product information
Options
- Integrated ventilation modes
- IPPV (Intermittent Positive Pressure Ventilation): Volume-controlled mandatory ventilation
- CPAP (Continuous Positive Airway Pressure)
- manual
- Other available options
- S-IPPV (Synchronized Intermittent Positive Pressure Ventilation): Assisted volume-controlled ventilation with a trigger time slot of 100 % of expiratory time.
- SIMV (Synchronized Intermittent Mandatory Ventilation): Assisted volume-controlled ventilation with a trigger time slot of 20 % of expiratory time.
- FlowCurve Pro: The FlowCurve Pro option includes flow measurement and Assisted Spontaneous Breathing (ASB) pressure support. This assists the patient’s respiratory effort.
- Pressure-controlled ventilation modes: The option Pressure-controlled ventilation modes with the ventilation modes PCV, aPCV, BiLevel + ASB and PRVC + ASB improves the supply of ventilated patients.
- CBRN (chemical, biological, radiological and nuclear hazards): The CBRN option together with the CBRN connection kit facilitates attending incidents under CBRN conditions. This enables ventilation to be performed in unconventional emergency situations.
Accessories & portable units
If you are still looking for the right accessories, then take a look at the catalog for our scope of supply! It contains the matching WEINMANN Emergency accessories and consumables for your MEDUVENT Standard.
Matching portable units
- LIFE-BASE light XS
- LIFE-BASE 1 NG XS
- LIFE-BASE 1 NG XL
- LIFE-BASE 3 NG
here you can find our portable units
Mobile transport solution
Download area
If you are looking for brochures, images, videos, PC simulations, instructions for use or certificates, you can find all product information in our Download area.
LEARN MORE
Technical details
- Dimensions (W x H x D):
206 mm x 137 mm x 130 mm
- Weight, incl. battery:
2.1 kg
- Product class according to Directive 93/42/EEC:
IIb
- Temperature (operating conditions):
-20 °C to +50 °C
- Humidity (operating conditions):
15 % rh to 95 % rh, no condensation
- Air pressure (operating conditions):
540 hPa to 1,100 hPa
- Altitude above mean sea level (operating conditions):
-500 m to 5,000 m
- Operating time (lithium ion battery):
average 7.5 hours
- Charging time (0 % – 95 %, lithium ion battery):
2.5 h (assuming settings for an adult)
- Display:
TFT color display 4.3"
- Data storage:
internal device memory and SD card
- Ventilation modes (mandatory/triggered):
- Volume-controlled:
IPPV, S-IPPV, SIMV (+ ASB) - Pressure-controlled:
BiLevel + ASB, PRVC + ASB, PCV, aPCV - Spontaneous:
CPAP (+ ASB) - Other:
Manual with MEDUtrigger
- Volume-controlled:
- Operating gas:
Ambient air; medical oxygen (100 % O₂) or concentrator oxygen (93 % O₂) can also be supplied
- Operating pressure range:
0.3 bar to 6 bar
- Monitoring:
- Measured values displayed: O2i, MVe, f(fsp), Vte, pPlat, pMean, pPeak, Vleak
- Bar graph: pressure with pAw
- Curves: airway pressure, flow
- Maximum outlet flow:
150 l/min
- Tidal volume:
50 ml to 2,000 ml
- Ventilation rate:
5/min to 40/min
- PEEP:
0 mbar to 20 mbar
- Pressure limit (pMax):
10 mbar to 65 mbar
- I:E:
4:1 - 1:4 / 1:1 in manual mode
- Standards used:
- Consistent serviceability confirmed on the ground and in the air to EN 13718-1, RTCA/DO 160G, EN 60601-1-12
- Ruggedness confirmed by successful shock and vibration tests to MIL-STD 810 G, RTCA/DO 160G and EN 1789
- Emergency and transport ventilator confirmed to EN 794-3 and ISO 10651-3

Ideal monitor layout
Displays the most important measurements and settings
User-friendly operation
Navigation knob is quick and easy to operate
Li-ion rechargeable battery
Average battery operating time of 8 hours (with typical ventilation settings for an adult patient)
Data storage and updates
SD memory card makes it easy to transfer device configurations and to perform software updates yourself
Connection for O₂ supply
Via O₂ cylinders, O₂ concentrators and gas supply systems
Connection for measuring circuit
Enables pressure to be measured and PEEP to be controlled, as well as oxygen to be routed to the patient
Accessories connection accessible from the front
For MEDUtrigger
Connection for ventilation hose
Connects the device to the breathing circuit
Hygiene filter
Protects the device from viral and bacterial contamination
Adapt oxygen supply flexibly

If you want to supply your patients with 100 % oxygen and then adapt the dose, simply connect the inhalation outlet of the pressure reducer to the oxygen inlet of MEDUVENT Standard using the globally-compatible oxygen inlet tube. You can then set the desired flow.
As MEDUVENT Standard runs purely on electricity, and not on compressed gas, the device does not use any oxygen to run. If you do not have any medical oxygen available but still want to ventilate with more than 21 % oxygen, you can use concentrator oxygen. To do this, just change the operating gas and connect the concentrator or the oxygen cylinder to MEDUVENT Standard.
It goes without saying that you can also operate MEDUVENT Standard with ambient air, without any oxygen source at all. The hygiene filter cleans the air used, removing viruses, bacteria, and particles, thus protecting both people and the device from contamination. You can select the CBRN (chemical, biological, radiological, and nuclear hazards) option to perform ventilation in unconventional situations.
Training courses and interactive training
Our training video gives you an initial impression of MEDUVENT Standard and explains how to operate it.
If you want to find out what it is like to operate one of the smallest turbine-driven ventilators in the world, you are welcome to use our interactive simulation software. You can use it to try out all the menus and functions of MEDUVENT Standard.